APPENDIX B – PROGRAMS
66
Report following an alarm
Model MILLENNIUM B2
S/N 03 BG 0001
Ver. SW Exxxx/BGyyyyyy
Counter 0007~0015
Selection 134c POROUS
Temperature 134 °C
Pressure 2.10 Bar
Process time 4 min
Stand-by HIGH
Pre-vacuum FRACTIONATED
Drying STANDARD
CYCLE START 19/11/02
11:30
Time T1 P T2 T3 T4
-------------------------------------------------------------------
00:01 CS 075.1 -0.00 130.9 115.2 093.4
00:11 074.9 -0.28 133.3 114.2 094.0
00:21 074.4 -0.46 146.3 113.2 094.5
00:31 074.3 -0.57 152.6 112.2 095.0
00:35 .. 074.3 -0.59 154.2 111.9 095.2
00:51 .. 078.9 -0.62 152.2 110.4 095.6
01:01 .. 074.9 -0.73 146.6 109.6 095.7
01:27 .. 047.8 -0.78 149.3 107.7 095.7
01:57 .. 047.8 -0.80 155.3 105.8 095.4
02:07 .. 076.5 -0.57 149.9 105.2 095.1
02:17 .. 081.1 -0.49 142.1 104.6 094.6
……………
……………
08:15 068.4 -0.76 151.8 104.7 102.3
08:22 … 061.1 -0.80 153.6 104.5 101.7
08:32 ... 097.4 +0.01 154.7 104.0 100.8
08:42 104.6 +0.24 148.9 103.7 101.0
……………
……………
15:04 135.5 +2.15 143.3 111.7 131.7
15:19 135.9 +2.17 148.5 113.5 132.6
15:28 135.3 +2.16 153.6 115.9 133.0
……………
……………
19:19 135.5 +2.15 157.4 126.5 132.5
19:34 134.4 +1.07 157.0 126.8 131.2
19:49 … 108.3 +0.25 156.4 126.8 119.9
19:53 DS 104.4 +0.00 156.1 126.6 116.2
STERILISATION NEGATIVE
ALARM CODE: A112
DESCRIPTION PTC SHORTCIRCUIT
CAUTION !
PLEASE REFER TO USER MANUAL
----------------------------------------------------------
Cycle Report
HELIX/BD TEST
Model MILLENNIUM B2
S/N 03 BG 0001
Ver. SW Exxxx/BGyyyyyy
Counter 0011/0019
Selection HELIX/BD TEST
Temperature 134 °C
Pressure 2.10 bar
Process time 3.5 min
CYCLE START 19/11/02
16:38
Time C bar
----------------------------------------------------------
00:01 CS 076.4 +0.00
02:06 1PV 089.3 -0.89
04:35 1PP 120.4 +0.99
05:45 2PV 062.5 -0.78
07:02 2PP 120.2 +0.97
08:15 3PV 061.1 -0.79
11:00 .. 135.6 +2.15
11:14 .. 136.0 +2.17
12:14 135.6 +2.14
13:14 135.6 +2.15
14:14 135.5 +2.14
14:45 .. 135.4 +2.14
15:20 .. 111.5 +0.00
16:34 … 047.8 -0.89
18:21 … 059.5 -0.86
19:21 .. 075.4 -0.50
20:06 CE 078.7 -0.04
12:33 MAX 136.0
14:44 MIN 135.4
Drying pulses 01
CYCLE END 19/11/02
16:38
HELIX/BD TEST COMPLETE
Please attach the indicator hereunder
OPERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
----------------------------------------------------------
Cycle Report
VACUUM TEST
Model MILLENNIUM B2
S/N 03 BG 0001
Ver. SW Exxxx/BGyyyyyy
Counter 0011/0019
Selection VACUUM TEST
CYCLE START 19/11/02
11:37
Time C bar
----------------------------------------------------------
00:00 CS 035.0 +0.00
01:39 E1F 037.4 -0.80
6:39 E2F 038.4 -0.79
16:39 E3F 042.0 -0.79
17:54 CE 045.5 -0.01
CYCLE END 19/11/02
11:41
VACUUM TEST: POSITIVE
OPERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
----------------------------------------------------------