APPENDIX E – ALARMS
82
CODE POSSIBLE CAUSE PROPOSED SOLUTION
Incorrect connection of the
temperature sensor (sterilization
chamber) to the connector.
A 111
Temperature sensor short circuit
(sterilization chamber).
Incorrect connection of the
temperature sensor (steam
generator) to the connector.
A 112
Temperature sensor short circuit
(steam generator).
Incorrect connection of the
temperature sensor (heating
element) to the connector.
A 113
Temperature sensor short circuit
(heating element).
Incorrect connection of the
temperature sensor (chamber wall)
to the connector.
A 114
Temperature sensor short circuit
(chamber wall).
Contact the Technical Support Department
(see Appendix Z
).
Intervention of the steam generator
safety thermostat.
Intervention of the heating element
safety thermostat.
A 200
Heating or steam generator heating
element malfunction.
Manually rearm the thermostat(s) located on the back of the device (see
the Chapter, Product Introduction”).
Unscrew the black plastic protection cap, press the red button
until you
hear a click and replace the cap.
Turn-off (RESET) and then turn-on the device.
If the problem persists contact the Technical Support Department
(see Appendix Z
).
Presence of water or condensate in
the sterilization chamber.
Perform RESET following the instructions.
Carefully dry the inside of the sterilization chamber and start the program
again.
Do not
put material impregnated with water, or liquids in general, in the
chamber.
Drain filter of the sterilization
chamber obstructed.
Clean or replace the drain filter
(See Appendix C
Maintenance”).
Air leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Vacuum pump broken.
A 250
Problem in the plumbing circuit.
Contact the Technical Support Department (see Appendix Z).
Water injection pump malfunction.
Problem in the plumbing circuit.
Contact the Technical Support Department (see Appendix Z).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 251
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z
).