APPENDIX E – ALARMS
84
CODE POSSIBLE CAUSE PROPOSED SOLUTION
Presence of water or condensate in
the sterilization chamber.
Perform RESET following the instructions.
Carefully dry the inside of the sterilization chamber and start the program
again.
Do not
put material impregnated with water, or liquids in general, in the
chamber.
Air leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Vacuum pump broken.
A 256
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z
).
Water injection pump malfunction.
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z
).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 257
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z
).
Steam leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with water
and start the program again.
Excessive load.
Perform RESET following the instructions.
Check the quantity of the material in the sterilization chamber and make
sure that it does not exceed the maximum allowed quantity, depending
on the type of load.
(See the Summary Table in Appendix A
, Technical Characteristics).
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 258
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z
).
Excessive load.
Perform RESET following the instructions.
Check the quantity of the material in the sterilization chamber and make
sure that it does not exceed the maximum allowed quantity, depending
on the type of load.
(See the Summary Table in Appendix A
, Technical Characteristics).
Steam leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with water
and start the program again.
A 259
Problem in the plumbing circuit.
A 260
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z
).
HAZARDS (category H)
H 150
Pressure sensor (MPX) broken.
Incorrect connection of the
pressure sensor (MPX) to the
connector.
H 160
Pressure sensor (MPX) short
circuit.
Contact the Technical Support Department
(see Appendix Z
).