Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
SERVICE MENU
39
SQ4 - COMPONENT TEST
Select
COMPONENT TEST to display:
- solenoid valves
- other components
- door
- tank management
SQ5 - SOLENOID VALVES
The status of the 6 solenoid valves of the manifold is
displayed and indicated with:
solenoid valve open
solenoid valve closed
During a COMPONENT TEST, with the door open, all
solenoid valves (EV) are powered off.
It is possible to check the opening / closing of
individual solenoid valves and the activation /
deactivation of radiator fans.
Several elements can be turned on at the same time to
check specific sections of the autoclave hydraulic
circuit.