During a COMPONENT TEST, with the door open, all
solenoid valves (EV) are powered off.
It is possible to check the opening / closing of
individual solenoid valves and the activation /
deactivation of radiator fans.
Several elements can be turned on at the same time to
check specific sections of the autoclave hydraulic
circuit.