Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
Select TEST CYCLES to display a vertical scroll menu with
the following options:
- No vacuum
- Continuous cycle
- Emptying cycle
Select NO VACUUM to display the cycle start-up screen.
With the command indicated on the left, you can
activate the cycle, which will be performed
without the pre-vacuum phase.
After the heating phase, the pressure rise is performed to
achieve the process values of 134°/2.10bar, a condition that
will be maintained for 4’.
This special cycle, which has no alarms, is designed to detect
any steam leaks from chamber, door, generator and
connected tubes.
Select CONTINUOUS CYCLE to display the cycle selection
screen.
Select and activate one of the cycles to run it and repeat it
automatically with a 5’ pause between subsequent cycles.