Chapter 6: Maintenance and Troubleshooting
Do only the maintenance described in this guide. Maintenance procedures other than those
specified in this guide must be done by qualified Molecular Devices personnel only. See
Obtain Support on page 124.
Before operating the instrument or doing maintenance operations, make sure that you are
familiar with the safety information in this guide. See Safety Information on page 5.
The following topics describe maintenance and troubleshooting procedures that can be
done by users to ensure optimal operation of the instrument.
Preventive Maintenance on page 104
Clean the Instrument on page 106
Before You Move the Instrument on page 106
Troubleshooting on page 123
WARNING! Service or maintenance procedures other than those specified in this
guide can be done only by Molecular Devices qualified personnel. When service is
required, contact Molecular Devices technical support.
Clean Injector and Accessories
CAUTION! Do not clean the inside of the injector other than the inside of the bottle
holder. Cleaning the inside can cause damage.
Periodically clean the outside surfaces, the inside and outside of the bottle holder, the
snorkel clamps, and the snorkel end of the injector tubing with a lint-free cloth that has been
lightly dampened with water. You can remove the bottle holder for cleaning. See Injector
Assembly and Maintenance on page 37.
If decontamination is required, use a lint-free cloth that has been lightly dampened with a
decontaminating solution, such as 70% ethanol or 3% sodium hypochlorite. See List of
Compatible Solutions on page 132.
Note: After you use a decontamination solution, always wipe the areas with a lint-free
cloth that has been lightly dampened with water to remove the residue. If you use
sodium hypochlorite, wipe the areas with a lint-free cloth that has been lightly
dampened with 70% alcohol before you wipe again with water.
CAUTION! Do not use abrasive cleaners. Do not spray cleaner directly onto the
instrument. Do not immerse the injector.
6

Table of Contents

Other manuals for Molecular Devices SpectraMax iD3

Questions and Answers: