Section 1, General Information
Chemgard Infrared Gas Monitor
Table 1-3.
Cross-Sensitivity Data for Ethylene Oxide
Cylinder P/N 80
4870, 100 ppm R-113 in Nitrogen,
available as a simulant span gas; see Calibratio
procedure for details
Calibrated 0-1000 PPM Ethylene Oxide in N
2
o
Table 1-6.
Cross-Sensitivity Data for Cyclo-Pentane
Cylin
der P/N 494450, 100 ppm Isobutylene in air,
available as a simulant span gas; see
Calibration procedure for details
GAS CONCENTRATION
2500 ppm 0
500 ppm 0
Calibrated 0-1000 PPM Cyclo-Pentane in N
2
o
Table 1-4.
Cross-Sensitivity Data for Ethyl Alcohol
Cylin
der P/N 804532, 0.75% Pentane in air,
available as a simulant span gas; see Calibration
procedure for details
GAS CONCENTRATION
ALCOHOL
Calibrated 0-1000 PPM Ethyl Alcohol in N
2
o
Table 1-2.
Cross-Sensitivity Data for Carbon Monoxide
Cylinder P/N 80
6734, 100 ppm CO in nitrogen,
available as a span gas; see Calibration
procedure for details
Table 1-5.
Cross
-Sensitivity Data for Hexane
Cylin
der P/N 10014894, 1000 ppm Pentane in air,
available as a simulant span gas; see
Calibration procedure for details
40 ppm 0
Calibrated 0-1000 PPM CO in N
2
Temperature: 25
o
C
Calibrated 0-1000 PPM Hexane in N
2
o