Chemgard Infrared Gas Monitor
Section 1, General Information
Table 1-19. Cross-Sensitivity Data for
Ethyl Ether
Cylin
der P/N 804868, 100 ppm R-22 in Nitrogen,
available as a simulant span gas; see
Calibration procedure for details
GAS CONCENTRATION
Calibrated 0-1000 PPM Ethyl Ether in N
2
o
Table 1-20. Cross-Sensitivity Data for
Dimethylamine
Cylin
der P/N 804868, 100 ppm R-22 in Nitrogen,
available as a simulant span gas; see
Calibration procedure for details
GAS
Calibrated 0-1000 PPM Dimethylamine in N
2
o
Table 1-22. Cross-Sensitivity Data for C4F6
Cylinder P/N 812784 30 ppm R123 in Nitrogen,
available as a simulant span gas; see Calibration
procedure for details
50 ppm 0
50 ppm 0
50 ppm 0
50 ppm 0
Calibrated 0-1000 PPM C4F6 in N
2
o
Table 1-18. Cross-Sensitivity Data for N-Pentane
Cylinder P/N 10014894 1000 ppm Pentane in
Air, available as a simulant span gas; see
Calibration procedure for details
Table 1-21. Cross
-Sensitivity Data for
Nitrogen Trifluoride
Cylinder P/N 804532 0.75% Pentane in Air,
available as a simulant span gas; see Calibration
procedure for details
GAS CONCENTRATION
N-PENTANE
GAS CONCENTRATION
TRIFLUORIDE
300 ppm 0
2500 ppm 0
500 ppm 0
Calibrated 0-1000 PPM N-Pentane in N
2
o
50 ppm 0
Calibrated 0-1000 PPM Nitrogen Trifluoride in N
2
o