MSA TECHNICAL DATA
ULTIMA
®
X
3 TM
23
7. Technical Data
7.1. Dimensions, weight
XE enclosure
Dimensions W x H x D (mm) 162 x 262 x 100
Weight approx. 5 kg
7.2. Performance Specifications
Gas types Combustible gases, oxygen and toxic gases
Operating range 0°C to 40°C
Extended range
*)
-20°C to +50°C
Operating range NH
3
*)
0°C to 30°C
Extended range
NH
3
, CL
2
, CLO
2
*)
-10°C to 40°C
Toxic gases and
oxygen
Calibrate within operating temperature the range.
Temperature
range
Combustible gas,
catalytic and IR
Local and remote sensor -40°C to +60°C-
Zero drift Less than 5% per year, typically
Span drift Less than 10% per year, typically
Noise Less than 1% FS
*) In the extended temperature range, the sensor may not meet all the specified performance
parameters.

Other manuals for MSA Ultima X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA Ultima X and is the answer not in the manual?

MSA Ultima X Specifications

General IconGeneral
Ingress ProtectionIP66/67
EnclosureExplosion-proof
Detection MethodElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, Toxic gases, Oxygen
DisplayLCD
Output4-20 mA, HART
Housing MaterialStainless steel
CertificationsATEX, IECEx, UL, CSA, SIL 2