EasyOne Pro™ / EasyOne Pro™ LAB 107
Hygiene, Cleaning, Maintenance, Disposal
Calibration Check with Test Subject
Some Basic Facts
Instead of a syringe a Biological Standard subject with known spirometry
values (BioCal Subject) performs a series of spirometry tests or DLCO tests.
These tests are observed and statistically controlled, if they are within an
expected range.
The data is collected on individual healthy, non-smoking BioCal Subjects
(e.g. physician, respiratory therapist, …).
Note
Also refer to www.ndd.ch. Application Note "DLCO Bio-
Calcheck EasyOne Pro".
Fig. 7-4 BioCal Check main screen
aehd f gb c
a Sensor selection
b Parameter value tabs
c Deselect tests (remove check mark)
d Click to conduct a new test
e Click to select the BioCal Subject
f Click to select a parameter
g Click to display the graphic trends
h Click to add a new BioCal Subject