Hygiene, Cleaning, Maintenance, Disposal
108 EasyOne Pro™ / EasyOne Pro™ LAB
Selecting a Test Subject and Performing the Calibration Check
Start the program.
Select the Utilities menu.
Open the Calibration Test menu and select Biological Calibration
Check.
The Biological Calibration Check menu (BioCal-Check main screen) is open.
Select the device according to its serial number a (check label on
sensor).
Select the BioCal Test Subject e. If the subject is not in the list, press Add
New BioCal Subject h and select the subject or add a new one.
Recommendation: Set the Patient ID to "Bio-1", "Bio-2", …
Afterwards return to the BioCal-Check main screen.
Click Add New Test d.
Perform the test (test selection: FVC, FVL, DLCO, Test-End, … according to
configuration settings)
Recommendation: Try to blow a QC grade "A" test.
Recommendation: Perform only one BioCal-Check test a day.
Press Finish Test.
To show the graphic trend, click the Graph button g.
To select another review parameter, change the selection in Displayed
Parameter Selection f.
If you selected a spirometry parameter, the spirometry BioCal quality
grades will be displayed. If you selected a DLCO parameter, the DLCO
BioCal quality grades will be displayed.
The parameter values are divided into two tab sheets b: Baseline Values
(used to calculate warning levels) and Control Values (controlled values for
BioCal-Check QC grade calculation).
To deselect a Test (Baseline and Control Values) remove the check mark c of
the corresponding test. A deselected test is not used for calculations any
more. This means also that by deselecting a baseline value, the first control
value is moved to the baseline values, because a minimum number (20) of
baseline tests is required for a valid analysis.