EasyManua.ls Logo

NDD easyone pro - DLCO Tab

NDD easyone pro
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EasyOne Pro™ / EasyOne Pro™ LAB 121
System Settings
DLCO Tab
On this tab, you set up the device for DLCO tests.
Warning
Inaccurate Test Results—Incorrect gas concentration entries
cause inaccurate test results.
Fig. 8-13 DLCO tab
a cb g hd e ij
f
a Selected gas concentration. Compare
the displayed values with the data on
the gas cylinder. The concentration
must be within the range indicated in
parentheses.
b Click to select the predicted values for
children
c Click to select the predicted values for
adults
d Select check box to automatically
determine the discard volume
e Click to select a parameter (see Fig.
8-10)
f Click to update the predicted values
(e.g., after selecting another predicted
value formula)
g System dead space
h Anatomic dead space
i Discarded volume
j Selected concentration of the simulator
test gas. Compare the displayed values
with the data on the gas cylinder. The
concentration must be within the range
indicated in parentheses.

Table of Contents

Related product manuals