EasyOne Pro™ / EasyOne Pro™ LAB 129
System Settings
Environment Ta b
On this tab you enter the ambient conditions. These will be overwritten by
values entered prior to the test (see "FVL Test (inspiratory and expiratory
measurement)" on page 51).
Fig. 8-22 Environment tab
a b
c
d eg h i
f
a Boxes for entry of room temperature,
relative humidity and altitude (above
sea level) of the examination site
b Measured values for room tempera-
ture, relative humidity, altitude (above
sea level), and case temperature
c BTPS conversion factor
d ATPS values will be used, yes/no (no
BTPS conversion)
e Select the check box, if you do not want
to enter ambient conditions for two
hours or until a restart. Otherwise,
remove the check mark by clicking
again.
f The values entered at a will be used for
BTPS calculation.
g Sensor case temperature
h Click to close the window, saving the
data
i Recalculate BTPS conversion factor
(e.g. after a change of the ambient con-
ditions)