EasyOne Pro™ / EasyOne Pro™ LAB 63
Performing a Test
Now proceed to the next test step (c, Fig. 3-25) and confirm with Done.
The next step will be displayed and the next action indicated at c is Wait 30
seconds.
When the waiting time is over, click Done (you can abort the waiting time
at any time with Done).
The next step will be displayed and the next action is Perform repeatable
FEV1 measurements.
Click Start to initiate the FEV1 measurement.
Note
After the FEV1 measurement, you need to check whether the
measurement quality is sufficient or whether additional
trials are required.
With key a (Fig. 3-24) you can display the test diagram
(response curve) at any time during the test.
Click Protocol to display the protocol screen: the measurement and the
next step will be displayed and the next action shown in this example is
Administer 0.0625 mg/mL concentration.
Fig. 3-26 Protocol screen