33
Setting
English
<Setting value>
Main menu Sub menu Setting details Setting range Factory setting
BLE SET PAIRING Pair the foot control with the control unit FC -
Pair Osseo 100+ with the control unit OS -
Pair the Link Module with the control unit LK -
Pair an external terminal with the control
unit
PC -
BLE OFF Unpair the foot control from the control
unit
FC -
Unpair Osseo 100+ from the control unit OS -
Unpair the Link Module from the control
unit
LK -
Unpair the external terminal from the
control unit
PC -
UNT SET BZ VOL Volume for beep that sounds at key
operations and when activating�
H: high
L: low
H
LCD BL Operation panel brightness 1 - 10
Larger value: Brighter
10
TS LV Touch sensor sensitivity level
(When sensitivity is poor such as when
using barrier film or surgical gloves)
1 - 3
Larger value: More
sensitive
1
CL LV1 Coolant solution flow
level
(per display)
1 - 100% 20
CL LV2
1 - 100% 40
CL LV3
1 - 100% 60
CL LV4
1 - 100% 80
CL LV5
1 - 100% 100
F-RESET Initializing settings
When display switches from off to on, press the enter key to restore settings to factory
settings�
FC SET BUT ST Foot control button guidance
(See “8 – 1 – 2 Foot control button guidance”)
See notes on
the left
PDL-SET Motor rotation specifications according to
amount the foot control speed control
pedal is depressed
VAR: Rotation speed changes within the
range set according to the amount
depressed
FIX: When depressed more than 50%, the
motor rotates at the set speed
VAR, FIX VAR
FC CAL Calibration of foot control pedal and buttons
(See “7 – 3 Foot control calibration”)
-