DMTA-10043-01EN, Rev. C, July 2016
Additional 27MG Gaging Features
47
Hold/Blank
•Measure rate
•CAL Lock
7.6.1 Changing Units
Changing units allows the user to change from inch measurement units to millimeter
measurement units.
To change units
1. Press [2
nd
F] [] (SETUP) to display the setup tabs.
2. Use the [, ] keys to highlight the MEAS tab.
3. Use the [] key to highlight UNITS and the [, ] keys to select imperial (IN) or
metric (MM) display units.
4. Press [MEAS] to return to measurement mode.
7.6.2 Changing Resolution
Resolution allows the user to change the number of decimal points displayed in the
thickness display.
The user can select between STANDARD (0.01 mm [0.001 in.]) and LOW (0.1 mm
[0.01 in.]).
To change resolution
1. Press [2
nd
F] [] (SETUP) to display the setup tabs.
2. Use the [, ] keys to highlight the MEAS tab.
3. Use the [] key to highlight RESOLUTION and the [, ] keys to select
STANDARD or LOW.
4. Press [MEAS] to return to measurement mode.
7.6.3 Configuring Min or Max Mode
The Min/Max mode allows the user to put the gage in minimum (MIN) or maximum
(MAX) scanning mode. This feature allows the user to scan over an area and quickly
determine the minimum or maximum thickness.