10-019116-01EN, Rev. 2, January 2021
Maintenance and Troubleshooting 49
4. Maintenance and Troubleshooting
This chapter contains maintenance procedures and some possible ways to resolve
issues that you may face during operation of the Vanta iX XRF analyzer.
4.1 Measurement Window Replacement
This section explains how to replace a Vanta iX measurement window. Olympus
recommends changing the window if it is dirty, contaminated, broken, or torn.
• Never conduct any test using a broken or torn window.
• Immediately replace any broken or torn window.
• For best results, regularly replace the window.
• Make sure the replacement window matches your Vanta iX series and method.
The correct window is required to achieve proper operation and accurate results.
4.1.1 Measurement Window Types
Table 7 on page 50 describes the measurement window types. You must specify the
proper film material to match your Vanta iX and your method (calibration).