Calibrating Open/Short/Load:
Open calibration.
Short calibration.
Note that the inductance of the short connection
is assumed to be zero. Short Delay Time is 0 s by
default.
Load calibration.
Default Load Resistor value setting is 50 Ω.
For more details and practical examples on how to perform calibration in the Voltage/Current
measurement mode, please refer to the corresponding application notes on
www.omicron-lab.com.
10.4 Further calibration information
10.4.1 Difference between Full-Range and User-Range calibration
Full-Range calibration and User-Range calibration differ only in the frequencies that are used to
measure the correction factors.
Full-Range calibration measures the correction factors over the "full" frequency range of the
instrument at factory-predefined frequencies.
User-Range calibration measures the correction factors at exactly the same frequency range and
frequency points that are used in the measurement currently configured by the user.