10.4.7 Exporting calibration data
User-Range and Full-Range calibration data can be exported to a .mcalx file for the use in the
Automation Interface.
To do so, click on the export icon in the File ribbon . This will open the export
pane where you can select the calibration export . The following picture
shows the export settings:
In the calibration export pane, you can select any available calibration to be exported. A .mcalx file can
contain more than one calibration. When loading in the Automation Interface, any usable calibration
will be loaded and activated if possible. Please make sure that the device settings in the Automation
Interface match the settings in the Bode Analyzer Suite.
The .mcalx file can be loaded only in the Automation Interface (see: 12 Automating
measurements on page 154 !
To save a calibration for use in the Bode Analyzer Suite, please simply save the
measurement file (.bode3) that contains the calibration as well.