EasyManua.ls Logo

Omicron Lab Bode 100 - Series-Thru

Omicron Lab Bode 100
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The following table shows some Rs values and their influence on the impedance measurement range:
Series Resistor Rs Lower Z Limit Upper Z Limit
200 Ω 5 mΩ 1125 Ω
450 Ω 10 mΩ 2250 Ω
499 Ω 11 mΩ 2480 Ω
950 Ω 20 mΩ 4500 Ω
The Shunt-Thru with series resistor configuration suffers from the same systematic error as
the normal Shunt-Thru configuration. Please refer to 8.2.3 Shunt-Thru on page 66 for more
details about the cable-braid-error.
Calibration
The Shunt-Thru with series resistor measurement requires at least one Thru calibration.
Thru calibration must be performed with the series resistors included.
For more details on calibrating a Shunt-Thru measurement, please refer to 10.3.4 Calibrating a Shunt-
Thru or Series-Thru measurement on page 100.
8.4.5 Series-Thru
The Series-Thru measurement configuration is especially suitable to measure very high impedance
values. It is basically an S21 transmission measurement in a special configuration. The impedance is
calculated from the measured S21 parameter.
Measurement information
Bode 100 measures S21 gain and calculates impedance Z using the equation:
The Series-Thru configuration offers high sensitivity for high-impedance DUTs. Impedance values in
the MΩ region can be measured. Using an output amplifier further increases the upper impedance
measurement limit.
Shunt-Thru measurement configuration is not suitable for low-impedance DUTs. It is not
recommended to measure impedance values below roughly 1 kΩ.
One advantage of the series-thru connection is that a shielded test-setup can easily be constructed.
Measurement types and applications
OMICRON Lab
69

Table of Contents

Other manuals for Omicron Lab Bode 100

Related product manuals