EasyManua.ls Logo

Omicron Lab Bode 100 - Calibrating a Reflection or One-Port Impedance measurement

Omicron Lab Bode 100
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10.3.1 Calibrating a Reflection or One-Port Impedance
measurement
In this section you learn how to calibrate an Impedance, Reflection or Admittance measurement in the
Transmission / Reflection measurement mode or in the One-Port impedance measurement
mode.
The Transmission / Reflection mode offers the possibility to measure Transmission and Reflection
sequentially. Please note that the Impedance (Open/Short/Load) calibration is only applied to a
Impedance, Reflection or Admittance measurement. If you want to calibrate a Gain measurement
please refer to 10.2.1 Calibrating a Transmission (S21) measurement (see page 88).
In the Transmission / Reflection measurement mode or in the One-Port impedance measurement
mode both receivers are internally connected to the 50 Ω source resistance. Bode 100 is internally
calibrated such that it measures the Impedance/Reflection directly at the OUTPUT port of the device.
The reference plane is directly at the OUTPUT port (BNC connector).
When connecting a DUT with a coaxial cable, the cable introduces additional impedance and time-
delay. To compensate these effects, the reference plane can be moved to the end of the cable by
performing an Open/Short/Load calibration at the end of the cable.
Figure 10-4: Shifting the reference plane by performing Open, Short and Load Calibration
94
OMICRON Lab

Table of Contents

Other manuals for Omicron Lab Bode 100

Related product manuals