10.3.3 Calibrating an Impedance Adapter measurement
In this section you learn how to calibrate an Impedance, Reflection or Admittance measurement in the
Impedance Adapter measurement mode.
In the External Coupler measurement mode and the External Bridge impedance measurement
mode both receivers are connected to the input Channel 1 and Channel 2 at the front panel of
Bode 100.
The Impedance Adapter measurement mode is designed for component impedance measurements
performed with the B-WIC and B-SMC impedance test-fixtures from OMICRON Lab.
B-WIC and B-SMC contain a resistive measurement bridge, which is specifically optimized for
Bode 100.
The Impedance Adapter measurement mode requires impedance calibration. You cannot
start a measurement without having Open, Short and Load calibration performed.
Figure 10-6: Calibrating the B-WIC impedance test fixture by performing Open, Short and Load
Calibration