The following pictures show how to perform Open, Short and Load calibration steps for the B-WIC and
B-SMC impedance test fixtures. Note that in the Impedance Adapter measurement mode the default
value for Load Resistor is 100 Ω and the default value for Short Delay Time is 0 ps.
Open (B-WIC) Open (B-SMC)
Short (B-WIC) Short (B-SMC)
Load (B-WIC) Load (B-SMC)
Calibration / Correction
OMICRON Lab