7.2.4 Status bar
The Status bar has several displays and interactive control elements. These are:
Source indicator
The source indicator shows if the signal source of Bode 100 is
switched on. By moving the mouse over the indicator, a pop-up
allows to change between Auto off and Always on setting. More
details regarding the source setting can be found in 9.4 Signal
source settings on page 80.
Overload indicators
The overload indicators show the signal level at the receivers of
Bode 100. Red color indicates an overload of the receiver. An
overload warning will be shown in the chart in such a case.
More details about the overload indicator can be found in
9.3 Choosing the receiver attenuator on page 78.
Internal device calibration indicator
Shows the date of the last performed internal device calibration.
Move the mouse over the indicator in order to perform a new
internal device calibration.
More details about the internal device calibration can be found in
10.1 Internal device calibration on page 85.
Device connection indicator
Displays the serial number of the connected device. If no device
is connected, "No device" is shown. Move the mouse over the
indicator in order to search or re-connect to a Bode 100 device.
Bode Analyzer Suite introduction
OMICRON Lab
29