To do so, perform Open, Short and Load calibration
as shown in the pictures below:
Connect Open and press Start.
Wait until the Open calibration has performed.
Check if your Short element fits to the Short
Delay Time Setting (50 ps default).
Connect Short and press Start.
Wait until the Short calibration has performed.
Check if your Load element fits the Load
Resistance Setting (50 Ω default).
Connect Load and press Start.
Wait until the Load calibration has performed.
Hint: The load delivered with the Bode 100 is marked with its exact impedance. You can
improve the calibration accuracy by entering this value in the Advanced Settings area.
The Short Delay Time in the Advanced Settings area has been chosen to match the Short
elements delivered with Bode 100. If your short is marked with Rosenberger you can set the
Short Delay Time to 40 ps, if it is marked with Radiall to 60 ps.