After you have copied the trace data to the clipboard, the Bode Analyzer Suite will show a "Clipboard
→ new memory" button on the right-hand lower corner:
Click the button to paste the trace data into any .bode3 file.
Bode Analyzer Suite does not check for data validity. You can mix Gain and Impedance data.
Please make sure to copy and paste the right data. If your measurement has Gain and
Impedance data, Bode Analyzer Suite will show a selection possibility which data to paste.
Paste / import data from a spreadsheet program
You can also import data from i.e. a spreadsheet program as a memory trace into the Bode Analyzer
Suite. To do so, you must copy data into the clipboard that can be interpreted as trace data. Please
consider the following rules:
The first column is always interpreted as the frequency in Hz.
The first line (header) will be skipped if it does not contain numerical data.
The second column is interpreted as the real part of the complex number.
The third column is interpreted as the imaginary part of the complex number.
The complex number is always interpreted as either a Gain value or as an Impedance value.
Use the same field separator and decimal separator as configured in the options! You can find
them under Options → Clipboard.
The following table shows an example of data that can be imported if the decimal separator is chosen
to be the dot and the field separator the tab.
Frequency Real Imaginary
1001.56 50.6 1e-6
10005.98 60.9999 100.89E-8
100700 0.777E02 0.00000125
You can also use five columns, then the interpretation follows the order Frequency, Gain, Impedance:
Frequency Gain Impedance
Real Imaginary Real Imaginary
data data data data data
There might be a situation where the amount of data in the clipboard does not fit the amount
of available measurements in a .bode3 file. In such a case, Bode Analyzer Suite will show a
selection possibility where you can choose to select Gain or Impedance data or choose to
paste data as Gain or Impedance.