Typical measurement times for a single point depending on the RBW setting are shown below:
Receiver bandwidth Typical single-point measurement time
1 Hz 2.8 s
3 Hz 940 ms
10 Hz 280 ms
30 Hz 95 ms
100 Hz 30 ms
300 Hz 10 ms
1 kHz 3.8 ms
3 kHz 2.1 ms
5 kHz 1.7 ms
Besides the RBW setting, the receiver attenuator plays an important role to improve the signal to noise
ratio (SNR).