121
The CT Analyzer Test Cards
6.7.3 Test Results
The following test results are displayed in the lower part of the Ratio card. In
addition to the results displayed in the Ratio card you can view the ratio and
phase table pages described in section 6.7.4 on page 122 and 6.7.5 on page
123.
Parameter Description
Ratio Current ratio error (in %) at the specified primary current (I
p
) and
burden.
Pol. OK: Polarity OK, phase angle is in the range of 0° ± 45°.
Failed: Wrong polarity of the CT or wrong polarity of the
measurement leads.
ε
C
Composite error in % at the specified primary current (I
p
) and
operating burden.
This parameter is only displayed if the IEC 60044-1 standard is
selected in the CT-Object card.
Phase Phase displacement (in minutes) at the specified primary
current (I
p
) and burden.
N Winding turns ratio.
εt
Turns ratio error acc. to IEC 60044-6 class TPS or IEC 60044-1
class PX.
RCF Ratio correction factor.
This parameter is only displayed if the IEEE C57.13 standard is
selected in the CT-Object card.
TCF Transformer correction factor.
This parameter is only displayed if the IEEE C57.13 standard is
selected in the CT-Object card.