CT Analyzer User Manual
166
10.6 Compact Flash Card Interface
10.7 Remote Control Interface
The remote control interface of the CT Analyzer is exclusively intended to
connect the CT Analyzer to a computer (e.g. running the CT Analyzer PC
Toolset software) or to the optional CT SB2 switch box (for multi-ratio CT
measurement).
As of serial number JHxxxx or newer, the CT Analyzer is equipped with a USB
interface and a RS232 interface.
Note: The user has to select the interface to be used in the CT Analyzer settings
before connecting the CT Analyzer to a computer (refer to section 2.3 on
page 23). The CT Analyzer will only communicate via the selected interface. It
will not be recognized by the computer if the CT Analyzer settings do not match
the interface used for connection.
10.7.1 RS232 Interface
The RS232 interface can be used to connect the CT Analyzer to a computer or
to the optional CT SB2 switch box.
Figure 10-1 Pin assignment for RS232 remote control interface
Compact Flash card interface
Card type CF type 1
Allowed memory size 16MB - 2GB
1nc
2 RxD (data in / receive)
3 TxD (data out / transmit)
4nc
5 GND (signal ground)
6nc
7 RTS (out)
8CTS (in)
9+ 5V
Do not connect!
For internal use only!
Housing: Shield (ground)
9-pole SUB-D connector, male
Figure shows outside view
onto the pins at the CT Analyzer!