49
Technical Data
7 Technical Data
7.1 Specifications
7.2 PC and CTA Interfaces
The PC interface of the CT SB2 switch box is exclusively intended to connect
the CT SB2 to a computer (e.g. running the CT Analyzer PC Toolset software).
The CTA interface of the CT SB2 switch box is exclusively intended to connect
the CT SB2 to a CT Analyzer.
Figure 7-1 PC interface on the CT SB2
Specifications
Mains connection Connector according to IEC 60320
Mains voltage 100 - 240V
AC
/ 50/60Hz / 0.2A
Mains fuses 2 x T2.0AH 250V,
(high-breaking capacity wire fuse 5 x 20mm)
Output voltage 0 - 120V
51
96
5 GND (signal ground)
4nc
3 RxD (data in / receive)
2 TxD (data out / transmit)
1nc
9nc
8 RTS (out)
7CTS (in)
6nc
PC interface
9-pole SUB-D connector, female
Outside view onto the sockets at the
CT SB2!