69
Operating and Configuring the CT Analyzer
5.7.2 Default Test Card Selection
Enabling or disabling test cards for the default test is done using the Default
Test Card Selection option from the Setting Menu. In this case, the test card
selection made in this page becomes effective when starting a new CT test.
If you want to enable or disable test cards for your currently active test only,
press the Select Cards soft key in the CT-Object card. In this case, the test card
selection made in this page is immediately effective after pressing the Back soft
key.
To enable or disable a test card, select its entry using the cursor keys
and then press the Add or Remove soft key. Enabled cards are marked with a
checkmark, see Figure 5-10.
Figure 5-10 Select Cards page for selecting the default test cards
Some cards require the existence of other cards. This means that if you are
adding such a card, the required other cards are automatically added, too. On
the other hand, if you remove a card which is required by another card, the other
card is removed, too. Figure 5-11 shows these dependencies.
Press the Main soft key in
any test card
Main Menu:
- Settings
Select soft key
Setting Menu:
- Default Test Card
Selection
Select soft key
-> Select Cards page
How to get there: