6.2 MR-Config. Card
Use the MR-Config. card to configure the multi-ratio CT test. In this card, set
the number of taps available on the CT,
the "tap in use", i.e., the tap combination (ratio) which is actually used and
connected to the operating burden during operation of the CT,
the nominal CT ratios for the tap combinations,
the nominal burden for each tap combination,
and select the tap combinations you want to test during the multi-ratio CT
test.
If necessary, use the cursor keys to scroll the table.
The current ratio I
pn
/I
sn
and the nominal burden for the full tap combination as
well as the operating burden have to be specified in the CT-Object card, see
section 6.1 on page 36.
Figure 6-3 MR-Config. card
The MR-Config. card can be switched to display
the tap combinations or intertap combinations of the CT with
the nominal burden values or the operating burden.
Refer to sections 6.2.1 and 6.2.2 on the following pages for more information.
MR-Config. card for a new multi-ratio CT
test after selecting a predefined multi-ratio
scheme for I
pn
in the CT-Object card.
MR-Config. card for a new multi-ratio CT
test after specifying the values for I
pn
, I
sn
and VA manually in the CT-Object card.
Fields underlined by dotted lines can be edited.