65
Operating and Configuring the CT Analyzer
5.7 Settings
In the Setting Menu page you can access the device settings or the default
settings for a new CT test. Select an entry using the cursor keys and
then press the Select soft key to open the corresponding settings page.
Figure 5-9 Setting Menu page
5.7.1 Available Options in the Settings Menu
Press the Main soft key in
any test card
Main Menu:
- Settings
Select soft key
-> Setting Menu page
How to get there:
Select Language Selection of the user interface language.
One user language can be installed on the CT Analyzer.
The English user interface is part of the firmware and
therefore always available. The second language can be
installed by the user as required (refer to section 5.8 on
page 73).
Select Startup
Mode
Using this option you can select the default test mode the
CT Analyzer comes up each time when it is switched on.
Select the mode using the soft keys:
Single Tap: The CT Analyzer automatically initializes a
normal single-ratio CT test.
Multi-Ratio: The CT Analyzer automatically initializes a
multi-ratio CT test using the CT SB2 switch box (see the
CT SB2 User Manual for more detailed information).
Quick: The CT Analyzer automatically initializes the
Quick Test measurement function (see chapter 8 on
page 135 for more detailed information).
Default Test Card
Selection
Select the test cards a new CT test should contain by
default. Please refer to section 5.7.2 on page 69.