CT SB2 User Manual
8
Safe Operation
Make sure to position the test equipment on dry, solid ground.
If working in other than laboratory environment only use the CT SB2 with a
solid connection to ground of at least 6mm
2
. Use a ground point as close as
possible to the test object.
Do not open the CT SB2. Opening the CT SB2 invalidates all warranty
claims.
Do not repair, modify, extend, or adapt the CT SB2 or any accessories.
Do not operate the CT SB2 under wet or moist conditions (condensation).
Do not operate the CT SB2 when explosive gas or vapors are present.
Use only original accessories available from OMICRON electronics GmbH.
Before putting the CT SB2 into operation, check it for visible damages.
Use the CT SB2 only in a technically sound condition and when its use is in
accordance with the safety regulations for the specific job site and
application.
Always be aware of the danger of the high voltages and currents associated
with this equipment. Pay attention to the information provided in this User
Manual.
When disconnecting cables, always start at the device feeding the power.
Never connect or disconnect a test object while the outputs of the connected
CT Analyzer are active.
If the CT SB2 seems to be functioning improperly, please call the OMICRON
electronics GmbH hotline (refer to chapter "OMICRON Service Centers" on
page 57).
For protection against parasitic currents or voltages, always connect the
equipotential connector on the CT SB2 to protective ground (PE). Only use
the original cable set supplied by OMICRON electronics GmbH.
Make sure that the terminals of the test object to be connected to the CT SB2
do not carry any voltage potential. During a test, the CT Analyzer (with the
CT SB2 connected) is the only permitted power source for the test object.
Only use wires with 4mm safety "banana" connectors and plastic housing for
connection to the front panel input/output sockets.

Table of Contents