CT Analyzer User Manual
168
10.8.2 Shock and Vibration
10.8.3 Mechanical Data
10.8.4 Safety Standards, Electromagnetic Compatibility (EMC)
Dynamics
Vibration Tested according to IEC 60068-2-6;
frequency range 10 … 150 Hz; acceleration
2g continuous (20 m/s
2
); 20 cycles per axis
Shock Tested according to IEC 60068-2-27
(operating mode); 15g / 11ms, half-sinusoid,
3 shocks in each axis
Weight, Dimensions and Protection
Weight < 8kg (17.6lbs) without accessories
Dimensions W x H x D 360 x 285 x 145mm
(14.2 x 11.2 x 5.7")
CE Conformity, Requirements
The product adheres to the specifications of the guidelines of the Council of
the European Community for meeting the requirements of the member states
regarding the electromagnetic compatibility (EMC) Directive 2004/108/EC and
the low-voltage Directive 2006/95/EC.
EMC
Emission
Europe
International
USA
EN 61326-1 Class A
IEC 61326-1 Class A
FCC Subpart B of Part 15 Class A
Immunity
Europe
International
EN 61326-1
IEC 61326-1
Certified Safety Standards
Europe
International
USA
EN 61010-1
IEC 61010-1
UL 61010-1