179
Error and Warning Messages
813.xxx Warning [813.xxx] One or more parameters could not be found. Some calculations may
not work.
Reason: The .xml report file read from the Compact Flash card is of a newer format than
the device can generate.
Solution: The loaded file is incompatible or the software version on your CT Analyzer is
older than the software version that has generated the report. To enable your
CT Analyzer to read this report, it is necessary to update the CT Analyzer
software.
814.xxx Warning [814.xxx] You tried to load an old report. One or more parameters could not
be found. Some calculations may not work.
Reason: One or more entries could not be found in the .xml report file.
Solution: The software version on the CT Analyzer is newer than the software version
used to generate the report. The report probably does not contain all
parameters supported by the new CT Analyzer software.
815.xxx Error [815.xxx] Folder must be empty. Can't remove folder <Folder Name>.
Reason: The folder selected for deletion is not empty. Only empty folders can be
deleted.
Solution: Navigate to the folder you want to delete. Open the folder and delete all
contained files and subfolders. Then you can delete the empty folder.
816.xxx Error [816.xxx] You tried to load a report built with a <Device>.
Reason: The .xml file seems to be not a CT Analyzer report.
Solution: The report cannot be loaded by the CT Analyzer.
817.xxx Error [817.xxx] The xml file has no valid OMICRON report style.
Reason: The .xml file has no valid OMICRON report style.
Solution: The report cannot be loaded by the CT Analyzer.
818.xxx Warning [818.xxx] Moving of folders not supported.
Reason: Moving of folders is not supported by the CT Analyzer.
Solution: Select only files to move, not folders. Using the shift key and the up/down
cursor keys it is possible to select any number of files within a folder.
819.xxx Warning [819.xxx] File name contains an invalid character.
Reason: An invalid character was used to specify the file name.
Invalid characters: \ / : * ? \ < > |
Solution: Only use valid characters for the file name.
820.xxx Error [820.xxx] Memory management error. Contact the next OMICRON service center.
Reason: Memory management error.
Solution: Switch the CT Analyzer off, wait a second then switch it on again. If this error
occurs more frequently, you should contact your next OMICRON service
center.
821.xxx Error [821.xxx] Could not update firmware.
Reason: An error has occurred in the boot loader software.
Solution: Insert a Compact Flash card with valid software (CTAnalyzer.bin) in the
directory A:\Omicron\ and switch the CT Analyzer off and on again.
823.xxx Error [823.xxx] Could not update user text!
Reason: An error has occurred in the user text loader software.
Solution: Insert a Compact Flash card with valid software (CTUser_xxx.bin) in the
directory A:\Omicron\ and try again.