Bode 100 User Manual
78
7.3 Calibration in the Gain/Phase Mode
(External Reference Connection – CH1)
To compensate for the cable and connection setup effects in the Gain/Phase
mode, proceed as follows:
1. Connect the Bode 100 and start the Bode Analyzer Suite.
Select the Gain/Phase mode.
2. Click the Device Configuration toolbar button to open the
Configuration window.
In the Configuration window, set the parameters for your measurement. In
our example we have chosen the following settings.