OMICRON 9
Safety instructions
Before performing tests using high voltage, observe the following instructions:
Do not use the test equipment without a good connection to substation ground.
Do not insert objects (for example screwdrivers, etc.) into any input/output socket.
Do not operate the CP TD1 under ambient conditions that exceed the temperature and humidity limits
listed in 12.2 "Environmental Conditions" on page 109.
Make sure to position the test equipment on dry, solid ground.
Do not operate the CP TD1 in the presence of explosives, gas or vapors.
Opening the CP TD1 invalidates all warranty claims.
Do not use an extension cable on a cable reel to prevent an overheating of the cord; run out the
extension cord.
►If the CP TD1 does not seem to function properly, do not use it anymore. Please call the OMICRON
technical support.
1.4 Handling cables
Always turn off the CP TD1 completely before you connect or disconnect any cable (disconnect the
CPC 100 from mains or press its Emergency Stop button).
The high-voltage cable must always be well attached and tightly connected to both the CP TD1 and
the test object. A loose or even falling off connector at the test object carrying high-voltage is life-
hazardous. Make sure the connectors are clean and dry before connecting.
At the CP TD1, press the high-voltage cable’s plug to the connector tightly and turn the screw cap
until you feel a mechanical stop. If you notice a rough-running of the screw-cap, clean the screw
thread and use a lubricant (Vaseline recommended).
At the test object, insert the high-voltage cables’ plugs carefully until you feel a "click" position. Now
they are locked. Confirm this by trying to pull them out. This should not be possible now.
Note: Tighten the plugs manually. Do not use any tools for that because that can damage the plugs
or connectors.
Insert the yellow banana plug (the high-voltage cable’s grounding) into the respective plug socket.
Do not connect any cable to the test object without a visible grounding of the test object.
The high-voltage cable is double-shielded and therefore safe. However, the last 50cm (20 inch) of
this cable have no shield. Therefore, during a test consider this cable a life wire and due to the high-
voltage life-hazardous!
When the CPC 100 is switched on, consider this part of the cable has to be in the high-voltage area
due to a hazard of electric shock!
Never remove any cables from the CP TD1 or the test object during a test.
Keep clear from zones in which high voltages may occur. Set up a barrier or establish similar
adequate means.
Both low-voltage measuring cables must always be well attached and tightly connected to the
CP TD1’s measuring inputs IN A and IN B.
Make sure to insert the red and blue marked cables into the corresponding measuring inputs: IN A =
red, IN B = blue.
Tighten the plugs by turning them until you feel a stop.
Note: Tighten the plugs manually. Do not use any tools for that because that can damage the plugs
or connectors.
Do not use any other cables than the ones supplied by OMICRON.

Table of Contents

Other manuals for Omicron CP TD1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CP TD1 and is the answer not in the manual?

Omicron CP TD1 Specifications

General IconGeneral
BrandOmicron
ModelCP TD1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals