7. Connect the air tube to the “AIR” outlet located on the instrument.
8. Reopen the irrigation clamp.
Note: If air bubbles are observed in the infusion set’s line or in the irrigation
line, ask the surgeon to activate just irrigation until the bubbles have been
expelled, before reinserting the surgical instrument’s tip in the patient’s eye.
U/S
Do not activate the phaco U/S handpiece with the phaco tip in air. The
ultrasonic power should be applied to the phaco handpiece with the phaco tip
immersed in a test chamber filled with infusion fluid or in a beaker of sterile
fluid at ambient temperature. Otherwise handpiece damage may result.
Do not check vibration by placing hand or finger against handpiece tip.
Prolonged exposure to or direct contact with the vibrating tip may cause
damage to healthy tissue.
Always use the lowest ultrasonic output power level which is compatible to
the surgical application.
The phaco handpiece must be connected to the PULSAR
2
irrigation/aspiration
system and must not be used without it.
Vit
Do not activate the vitrectomy handpiece with the tip in air. The tip should
always be test run in a beaker of sterile solution. Testing in air will cause
irreparable damage to the tip.
Footswitch
Never lift or move footswitch by the cable. Damage may result.
The footswitch is not autoclaveable.
Sterilization
The re-use of materials labelled as “For single use” may be cause of
malfunctions and/or infections which could severely affect patient’s safety.