9.3.2 ACCESSORIES
1) Avoid dropping or mishandling handpieces and accessories. It is critical that
these components be handled with outmost care and inspected thoroughly after
each use for any damage or wear.
2) Periodically lubricate all external "O" rings on handpieces, cannulas, I/A
handpiece, and connectors etc.. with silicone grease or oil. DO NOT OVER-
LUBRICATE.
3) Periodically inspect the fluidic lines, fittings, and handpieces for any damage or
wear.
4) Reassemble all parts before storing.
5) Place tip protective caps over all handpieces before wrapping and storing.
CAUTION:
1) Do not use synthetic detergents or oil-based soaps.
2) Be sure that the diathermy handpiece is completely dry before
using it, as RF currents may find a path through wet surfaces.
3) Vitrectomy handpieces should never be completely immersed in
distilled water as the water may enter the high pressure membrane
chamber thus causing malfunction.