9.2.2 SINGLE USE ACCESSORIES
1) Do not reuse microsurgical instruments which are labelled for single use.
9.2.3 REUSABLE ACCESSORIES
1) Reusable microsurgical instruments must be thoroughly cleansed before the
steam sterilization. When proceeding otherwise, residues can bake during steam
sterilization providing a protective layer preventing micro organisms being
destroyed effectively. Furthermore, any adhering contaminants might damage
plastic items in particular, due to the high temperatures. Do not subject any of
the microsurgical instruments to "chemclave" or to hot air sterilization.
2) For detailed information on cleaning, decontamination and sterilization of the
reusable surgical instruments, refer to the relevant instructions for use.
9.3 MAINTENANCE
9.3.1 UNIT
1)
The PULSAR
2
surgery system should be stored in a clean and dry environment at
room temperature.
2)
Remove I/A tubing after each use
.
3)
If water is found in the glass container of the gas filter (Fig.3, No.4) evacuate it
by depressing the pin valve (Fig.3, No.11) at the lower section of the filter.
4)
To maintain optimal performance through the full life of the equipment, a
yearly maintainance and calibration program should be established. Please,
inquire with your local Optikon Distributor/Service Center.
NOTE:
To maintain system safety, the compliance to EN60601-1 leakage current limits should be tested at
least once a year. Please, inquire with the Bioengineering Department in your clinic or with your
local Optikon Distributor/Service Center.