Installation and Operating Manual
OPTIKON 2000
Cod. 111007EN Rev.E
8-17
11) Depress footswitch beyond standby position to activate irrigation. Irrigation can
also be activated by the "CONTINUOUS IRRIGATION" button located on the
footswitch or by pressing the <IRR> soft key on the LCD screen.
12) Depress footswitch beyond mechanical detent #1 to activate IRRIGATION and
ASPIRATION. If in SURGEON mode, the footswitch will linearly control VACUUM
level. In this condition the footswitch depression from detent #1 to max travel
linearly varies the VACUUM level from 5 mmHg to the preset maximum level.
13) If footswitch is programmed for single linear mode, depress the main paddle
beyond mechanical detent #1 to activate vitrectomy (besides irrigation and
aspiration); rotate the main paddle to the right to disable vitrectomy. If in
SURGEON mode, the footswitch will linearly control CUT RATE level. In this
condition the footswitch depression from detent #1 to max travel linearly
varies the CUT RATE level from 60 cuts/min to the preset maximum level.
14)
If footswitch is programmed for double linear mode rotate the main paddle to
the right to activate vitrectomy. If in SURGEON mode, the footswitch will
linearly control CUT RATE level. In this condition, the main paddle rotation
linearly varies the CUT RATE level from 60 cuts/min to the preset maximum
level.
The vacuum level is automatically reduced to zero and a simultaneous VENTING (to
air or to BSS) occurs whenever the footswitch is released.
NOTE:
a. Semi-disposable vitrectome 123001 is reusable and autoclavable approximately 10 times.
b. Changing a parameter will not undo the default parameters. To recall default parameters or to
permanently store the new settings in the equipment memory, see paragraph "8-7. TO RECALL
OR STORE A PROGRAM".
CAUTION:
Vitreous cutter tip should always be tested in a beaker of sterile
solution. Testing in air will cause irreparable damage to the blade.
CAUTION:
Aspiration should always be tested in a beaker of sterile solution.
8.5 AUTOMATIC I.V. POLE
If the automatic I.V. pole is connected to the PULSAR
2
, the irrigating solution
bottle will move to the programmed position when the equipment is turned on.
The operator can adjust the bottle position using the UP/DOWN rocker footswitch
and/or the <UP>/<DOWN> soft keys of the Irrigation windows.