5-6
5.4 EMC TABLES
5.4.1 GUIDANCE AND MANUFACTURER’S DECLARATION ELECTROMAGNETIC
EMISSIONS
The Pulsar
2
ESP is intended for use in the electromagnetic environment specified
below. The customer or the user of the Pulsar
2
ESP
should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic Environment – Guidance
RF emissions
CISPR 11
Group 1
The Pulsar
2
ESP
uses RF energy only for its
internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Pulsar
2
ESP is suitable for use in all
establishments including domestic
establishments and those directly connected to
the public low voltage supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies