8-6
8.3 PHACOEMULSIFICATION
User Interface
Parameter windows available
- Irrigation
- Aspiration
- Ultrasound
Irrigation parameters
- Status (ON/OFF)
- I.V. Pole height
Aspiration parameters
- Status (Venturi or
Peristaltic)
- Vacuum mmHg (5 – 500)
- Flow rate cc/min (2-50,
peristaltic mode only)
- RiseTime
(slow/medium/fast)
- Mode (linear/panel)
Ultrasound parameters
- Phaco tip stroke m (5
100)
- Mode (linear/panel)
- U/S Rate
(Continuous,Short
Pulse,Single Burst, Multi
Burst, Continuous Burst,
Programmable Emission
Mode.)
- Pulse Frequency (1-40
bursts/sec)
- Timer (reading and reset
key)
-
Equivalent Phaco Time
-
AVG (Average Power
used)
Touching the U/S Function Key will display U/S Memory1 screen (U/S 1). In order to
display U/S Memory2 screen (U/S 2) touch the same key twice. Depressing the
same key a third time causes display of U/S Memory3 screen (U/S3); further
depression will reactivate U/S1. Swapping between memories U/S 1, U/S 2 and
U/S 3 can also be achieved by means of the footswitch. For footswitch operation
please refer to the relevant section.