PULSAR
2
ESP
Installation and Operating Manual
OPTIKON 2000
Cod. 111007EN Rev.E
5-2
5.2 TECHNICAL SPECIFICATIONS
PARAMETER SPECIFICATION
Manufacturer:....................... OPTIKON 2000 S.p.a.
via del Casale di Settebagni, 13
00138 Rome - Italy
Model: ................................ PULSAR
2
Regulatory compliance: .......... 93/42/EEC Medical Devices Directive (MDD)
Technical standards: .............. EN 60601-1 ; EN 60601-1-1 ; EN 60601-1-2 ; EN
60601-2-2
ENVIRONMENTAL SPECS
Storage: .............................. temp range –10°C to +70°C, humidity 10-100%
(non condensing)
Operation: ........................... temp range +10°C to +40°C, humidity 30-75%.
ELECTRICAL SPECS
Input voltage: ....................... 100/120/220/230-240 V A.C. select.
Frequency: .......................... 50/60 Hz
Power consumption: ............... 150 W
Line fuses: ........................... 220/240 Volt: 2,5 A T
100/120 Volt: 5 A T
COMPRESSED AIR SPECS
Input air pressure: ................. from 500 to 800 KPa (72 to 116 PSI)
Air consumption: ................... 30 normal litre/minute
IRRIGATION
Fluid delivery: ...................... gravity fed - eye pressure determined by the height of
irrigation source
Valving element: ................... solenoid driven pinch valve
Control: .............................. system footswitch
ASPIRATION
Aspiration pump types: ............ Venturi and peristaltic
Actuating medium: ................. pressurized air from external source
pressure: 500÷800 KPa (72÷116 PSI)