PULSAR
2
ESP
Installation and Operating Manual
OPTIKON 2000
Cod. 111007EN Rev.E
8-2
it is always advisable to begin with a low setting (such as 30%) increasing it in
case gradually up to obtain desired surgical effect. The bipolar diathermy can
operate both in fixed panel power control mode or linear surgeon power
control mode. In the latter case, the delivered power is linearly controlled by
the footswitch depression from 5% to the preset limit. To switch from LINEAR
SURGEON mode to FIXED PANEL or vice versa, use toggle <MODE> key in the
Diathermy setup pop-up window. The setup window will close down
automatically once you have effected the relevant adjustments.
NOTE:
Adjusting a parameter will not undo the program default parameters. To recall the default
parameters or to permanently store the new settings in the machine memory, see paragraph "8-7.
TO RECALL OR STORE A PROGRAM".
5) When diathermy is activated by depressing footswitch, diathermy power is
delivered to the handpiece and an audible tone of continuous pitch is
automatically generated to alert the operator.
NOTE:
The diathermy system is complemented by a comprehensive range of ancillary equipments including
bipolar pencil erasers, microforceps and various probes tailored to surgeon specific requirements.