PARAMETERS SPECIFICATION
Available flow range: .............. from 2 to 65cc/min (peristaltic only)
Vacuum rise time: ................. adjustable on 25 levels (0.5s - 12s)
Linear aspiration mode: .......... linear aspiration control, through system pedal, (for
vacuum and/or flow) from zero to maximum user
programmable
Safety device: ...................... vacuum sensor, measures the vacuum in the aspiration
line
Control: .............................. system pedal
POSTERIOR VITRECTOMY (R-Evo Smart CR only)
Handpiece type: .................... pneumatically powered guillotine cutter (VIT)
Cutting method: .................... reciprocating motion
Default cut rate: ................... user programmable
Available cut rate: ................. from 60 to 5000 cuts per minute (OptiVit)
from 120 to 10.000 cuts per minute (Twedge)
Port size: ............................ 0.5mm
Actuating medium: ................ pressurized air from an external source
Operating pressure: ............... 2,6 bar approx
Linear cutting mode: .............. linear control of the cut rate, controlled via system
footswitch.
Control: .............................. system pedal
ANTERIOR VITRECTOMY
Handpiece type: .................... pneumatically powered guillotine cutter
Cutting methods: .................. reciprocating motion
Default cut rate: ................... user programmable
Available cutting rate: ............ from 60 to 1000 cuts per minute (Revo Smart E)
from 60 to 5000 cuts per minute (Revo Smart CR and S)
Port size: ............................ 0.5mm
Operating Modes: .................. pressurized air from an external source (R-Evo Smart S
and CR) or integrated dry compressor (R-Evo Smart E)
Operating pressure: ............... from 2.6 bar
Linear cutting mode: .............. linear cut rate from 0 to preset controlled via system
pedal
Control: .............................. system pedal
DIATH (DIATHERMIA)
Type: ................................. bipolar generator, the generator stops when the RF power
supply is not needed