29
PARAMETERS SPECIFICATION
Operating frequency: ............. 2 MHz
Rated power: ....................... 9W (200 Ohm LOAD)
Maximum voltage without load: . 100 V
Default power: ..................... user programmable
Available power: ................... from 5 to 100% (percent)
Linear mode: ....................... allows the linear control of the DIATHERMY power by
means of system pedal
Handpiece type: .................... bipolar microforceps, slim stat pencil eraser, handpieces
for intraocular diathermy
Diathermy cable: ................... two conductors, 26 Gauge, 75 Ohm, 200V max, steam
autoclavable; use only the original diathermy cable.
Control: .............................. system pedal
ILLUMINATION (R-Evo Smart CR only)
Source type: ....................... Two independent LED lamps
Luminous flux: ..................... 400+400lm
Intensity adjustment: ............ 20 levels + 4 filters for protection and improvement of
visualization of epiretinal membranes
AIR INJECTION SECTION (R-Evo Smart CR only)
Nominal pressure:.................. from 5 to 120 mmHg
Actual IOP: .......................... Nominal pressure ± 3 mmHg
Safety devices: ..................... Air activation sound
SILICONE INJECTION SECTION (R-Evo Smart CR only)
Nominal pressure: ................. from 0.4 to 5 bar
Signals: .............................. Silicone injection activation sound, Alarm for Low
pressure at air inlet
PHACOEMULSIFIER
Handpiece type: .................... piezoelectric available with four or six crystals
Frequency: .......................... approx. 40KHz
Tip stroke: .......................... from 0 to 100µm
Power control: ...................... panel or linear control of U/S power by means of system
pedal.
U/S mode: ........................... Linear or panel; continuous, Autolimit, short pulse, HD
pulse, single burst, multiple burst, continuous burst
Timer U/S: .......................... from 0.00 minutes to 9.59 minutes - Equivalent Phaco
Time display