EasyManua.ls Logo

OPTIKON R-Evo Smart E - Page 36

OPTIKON R-Evo Smart E
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
33
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The R-Evo Smart systems are intended for use in the electromagnetic environment specified
below. The customer or the user of the R-Evo Smart should assure that it is used in such an
environment.
Portable and mobile RF communications equipment should be used no closer to any part of
the R-Evo Smart, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter
Immunity
Test
IEC 60601-1-2
Test level
Complia
nce
level
Recommended separation distance
Conducted RF
EN 61000-4-6
3 Veff
150kHz to 80MHz
3 Veff
d= 30cm
Radiated RF
EN 61000-4-3
3 V/m
80MHz to 2.7GHz
3 V/m
d= 30cm
Immunity to
proximity fields
from wireless
RF
communication
devices
IEC 61000-4-3
TETRA 400
380 390 MHz
27 V/m
27 V/m
d= 30cm
GMRS 460
FRS 460
430 170 MHz
28 V/m
28 V/m
LTE Band 13, 17
704 787 MHz
9 V/m
9 V/m
GSM 800/900,
TETRA 800, iDEN
820, CDMA 850, LTE
Band 5
800 960 MHz
28 V/m
28 V/m
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 5
1700 1990 MHz
28 V/m
28 V/m
Bluetooth, WLAN,
802.11 b/g/n, RIFD
2450, LTE Band 70
2400 2570 MHz
28 V/m
28 V/m
WLAN 802.11 a/n
5100 5800 MHz
9 V/m
9 V/m
Interference may occur in the vicinity of
equipment marked with the following symbol:

Related product manuals