– If one of the locking tabs is pressed:
The trolley can be placed easily and effortlessly,
in any orientation, in the surgical room near the
operating table.
All– Locking tabs have been pressed in:
The tool carrier is secured in place and avoids any
movement. After placing the Medical Device in the
desired place in the operating area, press down on the
locking tabs.
Installation of the Medical Device in the operating room
To install the device in the operating room using the
optional cart, proceed as follows:
• Release all the locking tabs of the wheels.
• Grab the tool carriage from the handle and gently push it
towards the desired position.
• The trolley must be placed on a levelled floor.
• Press at least two of the locking tabs on the wheels and
make sure that the carriage can not move by itself.
• The power plug is used as a means of electrically
isolating all poles simultaneously from the MAINS
SUPPLY. Do not place the appliance in such a way that it
is difficult to disconnect the power plug from the wall
socket